ACTIVA PC NEUROSTIMULATOR

The Activa® PC neurostimulator from Medtronic is a dual-channel device capable of delivering bilateral stimulation with a single device.

Activa PC contains a nonrechargeable battery and microelectronic circuitry to deliver a controlled electrical pulse to precisely targeted areas of the brain. The device is typically implanted subcutaneously near the clavicle, connected to an extension and leads, which are implanted in the brain.

The Activa portfolio provides the ability to control the stimulation field with innovative interleaved pulses and patient-specific therapy groups. The Activa system utilizes internal memory to store information that can be accessed from the N'Vision® clinician programmer. The patient's medical history, therapeutic window parameters, and valuable information about actual therapy usage can be stored within the device. The sorting capabilities of the device help clinicians choose settings that will allow them to program a positive therapeutic response while maximizing power-usage efficiency.

Deep brain stimulation using the Activa PC neurostimulator is approved for the treatment of symptoms due to Parkinson's disease, essential tremor, and dystonia.


Troubleshooting of deep brain stimulators (Activa SC/PC/RC Medtronic PLC) sometimes results in a decision to replace a tunneled stretch-coil extension cable.

Henderson et al., present a simple technique to accomplish this atraumatically without a tunneling tool.

In the treatment of patients with deep brain stimulators, complication avoidance and efficiency of operative time are paramount. They sought to find the most safe, effective, and rapid method for performing the conceptually-simple yet technically-nuanced act of replacing lead extension cables.

#6 (8.0 metric) surgical steel 18“ (45cm) monofilament (Ethicon US, LLC), also known as #6 sternal wire, was connected in-line with DBS extension cables (Medtronic DBS Extension 37086-60) in novel fashion to overcome intra-procedural hurdles encountered during the past decade in a busy functional neurosurgery service.

Patients tolerate the procedure well and return home shortly after recovery with no complications.

A less expensive and faster technique for passing pulse generator extension cables may be the use of a sternal wire. Using the described technique, pulse generators may be quickly and safely adjusted from side-to-side and site-to-site as the clinical situation dictates 1).

1)
Henderson F Jr, Takacs I. Rapid and Minimally-Traumatic Replacement of Stimulator Extension Cables: Technical Note on a Novel Use for Sternal Wire. World Neurosurg. 2016 Oct 18. pii: S1878-8750(16)31041-5. doi: 10.1016/j.wneu.2016.10.053. PubMed PMID: 27769951.