aneurysm_clip

Aneurysm clip

Aneurysm clips and MRI

MRI concerns in patients with a cerebral aneurysm clip:

1. the danger of the MRI magnetic field causing the aneurysm clip to be pulled or torqued off of the aneurysm or to tear the neck

2. the artifact produced by the metal of the clip in the magnetic field

3. heat generated in the region of the clip: not clinically significant

The more ferromagnetic the clip, the larger the force exerted on it by the magnetic field and the greater the image distortion near the clip.

Stainless steel (SS) is classified as martensitic (ferromagnetic) or austenitic (non-ferromagnetic).

Cobalt-based superalloys are non-ferromagnetic and include Elgiloy (Sugita aneurysm clips), Phynox (Yasargil) and Vari-Angle (McFadden). Essentially all modern aneurysm clips are MRI compatible, but patients clipped before the 1990s may have ferromagnetic clips.

If in doubt at the time of aneurysm surgery, apply the following simple test: non- ferromagnetic clips cannot be lifted or dragged with a small magnet.

see Yasargil aneurysm clips.

Aesculap’s titanium aneurysm clips

Catalog

Yasargil Special clips

Come in a variety of patterns and styles. All of which have the following features:

Non-ferromagnetic, MRI-safe up to 3.0 Tesla

Minor artifacts in CT and MRI


Titanium aneurysm clips reduced MR artifacts by approximately 60% compared to stainless steel clips. Artifacts were further reduced by using SE-based sequences with a high bandwidth or, if necessary, GE sequences with a low TE 1). Titanium Standard Aneurysm Clips

Titanium Fenestrated Aneurysm Clips

Titanium T-Bar Aneurysm Clips

Titanium Reinforcing Aneurysm Clips

Titanium Mini Aneurysm Clips

Titanium Applying Forceps

Titanium Removing Forceps

- See more at: https://www.aesculapusa.com/products/neurosurgery/aneurysm-clips/yasargil-titanium#sthash.keZ9fDz3.dpuf

1)
Burtscher IM, Owman T, Romner B, Ståhlberg F, Holtås S. Aneurysm clip MR artifacts. Titanium versus stainless steel and influence of imaging parameters. Acta Radiol. 1998 Jan;39(1):70-6. PubMed PMID: 9498874.
aneurysm_clip.txt · Last modified: 2019/05/19 20:56 by administrador